How to setup the
scope (& be happy)
CHANNEL SETUP
-- Enable the needed input channels using the "CH<X>" buttons.
The buttons corresponding to enabled will light up
-- Select the time scale with the "SCALE" knob. The time scale defines
the time resolution. Both values (ns/division and ps/point) are shown
in the trigger panel drawn in left bottom corner of the screen.
TRIGGER SETUP
-- Select "EXT" (external) trigger source using the arrow buttons
located below the "SOURCE" label. The EXT trigger input is the second
BNC connector from the left ("AUX IN")
-- Select "NEG" (negative) slope with the button located below the
"SLOPE" label
-- Select "NORM" mode with the button located below the "MODE" label
-- Select "EDGE" mode with the button labeled "EDGE"
-- Set the threshold level knob labeled "LEVEL". The level shall be set
to -150mV. The current value is normally shown in the trigger panel
drawn in left bottom corner of the screen.
ACQUISITION SETUP
-- From the top menu on the screen select Horz/Acq --> FastFrame
Setup. A panel will appear in the bottom part of the screen
-- Switch ON the FastFrame mode with the button located below the
"FastFrame" label
-- Select the record length. This defines the number of point acquired
for each waveform. Alternatively, it sets the acquisition period
(record length * ns/division). Clink on the "Rec Length" field, two
panels will appear on the top right corner of the screen. Change the
record length with the knob on its right. A precise value can be set
double-clicking on "Rec Length" field
--Maximize the "Frame count". This defines the number of the event to
be stored in memory before to stop the acquisition. We always want to
have the maximum number here: this however depends on the record length
and on the number of enabled channels. Change the "Frame count" in the
same way of the "Rec Length"
-- On the right side, select "Horiz/Acq Setup". A new panel will open.
-- Select the tab "Acquisition".
-- Acquisition Mode --> Sample
-- Fast Acquisition --> Off
-- Roll mode --> Off
-- Sampling Mode --> Real Time Only
-- Select the "Horizontal" Tab. Here one can switch on the "Delay
mode". This is needed if the trigger and the signal are very far apart
in time. Refer to the drawing on the screen.
MOUNT SHARED DISK FROM THE SCOPE
minimize scope window (File --> Minimize)
From windows "standard" screen:
File --> My computer --> My network places --> Tools
--> Map network drive
B: ---> \\192.168.1.1\public
click on "different username" and use username dreamtest (usual
passwrd)
W. Vandelli, R. Ferrari